CTG2100F COATING THICKNESS GAUGE
 |
This equipment is a portable measuring instrument that can quickly and without injury, sophisticated manner Tu, coating thickness gauge. Can be used laboratory and the construction site.
Scope
The equipment used magnetic thickness measurement method, non-destructive measurement of magnetic metal matrix (such as steel, iron, steel alloys and hard-magnetic, etc.) on non-magnetic coating thickness (such as aluminum, chromium, copper, enamel, rubber, paint, etc. ).
This equipment can be widely used in electroplating, anti-corrosion, aerospace, chemicals, automobile, shipbuilding, light industry, commodity inspection and other testing areas. Professional materials is necessary to protect the equipment. |
Main features:
Calibration of two kinds: one point calibration,two point calibration
Display resolution: 0.1 μ m (measuring range of less than 100 μ m)
1μ m (range greater than 100 μ m)
Have five statistics, data storage 500
There are two working methods: direct ways and means of group
There are two measurements: continuous measurement and a single measurement
There are two shutdown: manual and automatic shutdown shutdown
Can be set Bound: The gauge of the measured value can be automatic alarm and a number of measurements available on the histogram value analysis;
Deleted features: the gross error and error settings can be deleted;
A music tone in the operation carried out at any time tips
A power supply under-voltage direct function
An error function
Main Specifications:
The use of ambient temperature: 0 ~ 40 ℃
Power: nickel-metal hydride batteries 9 V
Appearance size: 150 × 80 × 30mm
Weight: 300 g Range: 0 ~ 1500 μ m measurement error ------- ± (1% H +1): |
Configuration:
main frame 1
Substrate 1
Calibration foil 5
9V Battery 1
Probe F1 1
Manual Instruction 1 |
CTG2100N COATING THICKNESS GAUGE
 |
This equipment is a portable measuring instrument that can quickly and without injury, sophisticated manner Tu, coating thickness gauge. Can be used laboratory, construction site.
Scope
The equipment used swirl thickness measurement method, non-magnetic metal matrix (such as copper, aluminum, zinc, tin, etc.) on non-conductive coating thickness (such as: enamel, rubber, paints, plastics, etc.). This equipment can be widely used in electroplating, anti-corrosion, aerospace, chemical, automobile, shipbuilding, light industry, commodity inspection and other testing areas. Professional materials is necessary to protect the equipment.
|
Main features:
Calibration of two kinds: one point calibration, two point calibration
Display resolution: 0.1 μ m (measuring range of less than 100 μ m)
1μ m (range greater than 100 μ m)
Have five statistics, data storage 500
There are two working methods: direct ways and means of group
There are two measurements: continuous measurement and a single measurement
There are two shutdown: manual and automatic shutdown shutdown.
Can be set Bound: The gauge of the measured value can be automatic alarm and a number of measurements available on the histogram value analysis;
Deleted features: the gross error and error settings can be deleted;
A music tone in the operation carried out at any time tips
A power supply under-voltage direct function
Main Specifications:
The use of ambient temperature: 0 ~ 40 ℃
Power: nickel-metal hydride batteries 9V
size: 150 × 80 × 30mm
Weight: 300 g
Range: 0 ~ 1500 μ m (2% H +1) ± measurement error:
|
Configuration:
main frame 1
Substrate 1
Calibration foil 5
9V Battery 1
Probe N1 1
Manual Instruction 1 |
CTG3100FN COATING THICKNESS GAUGE
 |
This equipment is a portable measuring instrument that can quickly and without injury, sophisticated manner Tu, coating thickness gauge. Can be used laboratory and the construction site. Through the use of different probe, to meet the needs of a variety of measurements.
Scope
The equipment used magnetic and eddy current two thickness measurement method, non-destructive measurement of magnetic metal matrix (such as steel, iron, steel alloys and hard-magnetic, etc.) on non-magnetic coating thickness (such as aluminum, chromium, copper, enamel, Rubber, paint, etc.) and non-magnetic metal matrix (such as copper, aluminum, zinc, tin, etc.) on non-conductive coating thickness (such as: enamel, rubber, paints, plastics, etc.).
This equipment can be widely used in electroplating, anti-corrosion, aerospace, chemical, automobile, shipbuilding, light industry, commodity inspection and other testing areas. Professional materials is necessary to protect the equipment.
|
Main features:
Can use various probe (F400, F1, F1/90 °, F10, N1, N400, etc.) measurement;
Three calibration methods: one point calibration, two point calibration, the basic calibration;
Display resolution: 0.1 μ m (measuring range of less than 100 μ m)
1μ m (range greater than 100 μ m)
Have five statistics, data storage 500
There are two working methods: direct ways and means of group
There are two measurements: continuous measurement and a single measurement
There are two shutdown: manual and automatic shutdown shutdown
Can be set Bound: The gauge of the measured value can be automatic alarm and a number of measurements available on the histogram value analysis;
Deleted features: the gross error and error settings can be deleted;
A music tone in the operation carried out at any time tips
A power supply under-voltage direct function
An error function
Measurement error:
F1.N1.F1/90-------±( 2%H+1) two point calibration precision 1%
F10----±(2%H+10) two point calibration precision 1%
F400.N400-----±( 2%H+0.7) two point calibration precision 1%
Note: H - nominal value
Main Specifications:
The use of ambient temperature: 0 ~ 40 ℃
Power: nickel-metal hydride batteries 9V
Appearance size: 150 × 80 × 30mm
Weight: 300 g
Measuring range:
F1、N1、F1/90----0---1500um
F400、N400----0-400umF10: 0---10mm
|
Basic configuration:
main frame 1
Substrate 1
Calibration foil 5
9V Battery 1
Probe F1/N1 1
Manual Instruction 1 |
CTG4100FN COATING THICKNESS GAUGE
 |
This equipment is a portable measuring instrument that can quickly and without injury, sophisticated manner Tu, coating thickness measurement. Can be used laboratory and the construction site. Through the use of different probe, to meet the needs of a variety of measurements.
Scope
The equipment used magnetic and eddy current two thickness measurement method, non-destructive measurement of magnetic metal matrix (such as steel, iron, steel alloys and hard-magnetic, etc.) on non-magnetic coating thickness (such as aluminum, chromium, copper, enamel, Rubber, paint, etc.) and non-magnetic metal matrix (such as copper, aluminum, zinc, tin, etc.) on non-conductive coating thickness (such as: enamel, rubber, paints, plastics, etc.).
This equipment can be widely used in electroplating, anti-corrosion, aerospace, chemical, automobile, shipbuilding, light industry, commodity inspection and other testing areas. Professional materials is necessary to protect the equipment.
|
Main features:
Can use various probe (F400, F1, F1/90 °, F10, N1, N400, etc.) measurement;
Three calibration methods: one point calibration, two point calibration, the basic calibration;
Display resolution: 0.1 μ m (measuring range of less than 100 μ m)
1μ m (range greater than 100 μ m)
Have five statistics, data storage 500
There are two working methods: direct ways and means of group
There are two measurements: continuous measurement and a single measurement
There are two shutdown: manual and automatic shutdown shutdown
Can be set Bound: The gauge of the measured value can be automatic alarm and a number of measurements available on the histogram value analysis;
Deleted features: the gross error and error settings can be deleted;
Printing: Print Measurement measurement, statistics, gauges, histogram
A music tone in the operation carried out at any time tips
A power supply under-voltage direct function
An error function
And printers, computer communications (communications software operating environment for the Window operating system) connectivity.
Main Specifications:
The use of ambient temperature: 0 ~ 40 ℃
Power: nickel-metal hydride batteries 9V
Appearance size: 150 × 80 × 30mm
Weight: 300 g
Measuring range:
F1.N1.F1/90 0—1500um
F400.N400---0-400um
F10.N10 0-10um
Measurement error:
F1.N1.F1/90-----±(2%H+1) two point calibration precision1%
F400.N400----±(2%H+0.7) two point calibration precision1% F10.N10-------±(2%H+10) two point calibration precision1%
Note: H - nominal value |
Configuration:
main frame 1
Substrate 1
Calibration foil 5
9V Battery 1
Probe F1/N1 1
Manual Instruction 1
CTG4100 printer 1 |
CTG5100FN Coating thickness gauge

2 Measuring modes: continuous/single
2 Shutdown modes: manual/automatic
Wide measuring range with 5 probes available (next page)
Direct testing mode and block statistics mode (APPL/BATCH)
Direct print out of statistical values
Dataview to connect with PC
500 data can be stored
Technical specifications

Standard delivery
Main unit |
1 |
Probe F1 or N1 |
1 |
Calibration foils |
1 |
Instruction manual |
1 |
Cetificate |
1 |
Warranty card |
1 |
Carrying case |
1 |
Communication cable |
1 |
PC software Dataview |
1 |
Optional accessories
5 probes for different applications
Calibration foils in various thickness

Option probes for CTG510FN
technical parameters of probe
F-use magnetic probe, based on measurable magnetic metal (such as steel, iron, steel alloys and hard-magnetic, etc.) on non-magnetic coating thickness (such as zinc, aluminum, chromium, copper, rubber, paint, etc.).
N-type probe use of eddy current way, measurable magnetic non-ferrous metal matrix (such as copper, aluminum, zinc, tin, etc.) and non-conductive, Austenitic stainless steel cover the thickness (such as: rubber, paints, plastics, anodizing film ).
F / N-probe (except N400) gage with the times and can be used, other brands gage would need change.
Type F probe :
Probe |
F400 |
F1 |
F1/90° |
F10 |
Measuring Principle |
Magnetic method |
Measuring range(mm) |
0~400 |
0~1250 |
0~10000 |
Min resolution(mm) |
0.1 |
0.1 |
10 |
Tolerance |
One-point calibration (mm) |
±(2%H+0.7) |
±(2%H+1) |
±(2%H+10) |
Two-point calibration (mm) |
±(1%H+0.7) |
±((1%H+1) |
±(1%H+10) |
Minimum radius of curvature (convex) (mm) |
1 |
1.5 |
flat |
10 |
Minimum measuring area(mm) |
F3 |
F7 |
F7 |
F40 |
Minimum thickness of base material(mm) |
0.2 |
0.5 |
0.5 |
2 |
Type N probe :
Probe |
N400 |
N1 |
CN02 |
Measuring Principle |
Eddy current method |
Measuring range (mm) |
0~400 |
0~1250 |
10~200 |
Min resolution(mm) |
0.1 |
0.1 |
1 |
Tolerance |
One-point calibration (mm) |
±(2%H+0.7) |
±(2%H+1.5) |
±(2%H+1) |
Two-point calibration (mm) |
±(1%H+0.7) |
±(1%H+1.5) |
-------- |
Minimum radius of curvature (convex) (mm) |
1.5 |
3 |
flat |
Minimum measuring area(mm) |
F4 |
F5 |
F7 |
Minimum thickness of base material(mm) |
0.3 |
0.3 |
0 |
Note:H——Measured Value
Coating Thickness Gauge CTG8829
|
Description:
Operating principle: magnetic induction/eddy current (F/NF)
With separate probe(s)
Measuring range: 0-1250um/0-50mil
Resolution: 0.1/1
Accuracy: ±1-3%n or ±2.5um
Min. measuring area: 5mm
Min. sample thickness: 0.3mm
Battery indicator: low battery indicator
Metric/ imperial: convertible
Power supply: 4x1.5V AAA(UM-4)battery
Auto power off
Operating conditions: 0-+45?(32?-104?),=90%RH
Dimensions: 124x62x30mm
weight: 115g |
Standard Delivery of TG8829FN
Main unit 1
F type probe 1
N type probe 1
F calibration base set 1
N calibration base set 1
Calibration foil set 1set (4 foils)
Carrying case 1
Instruction manual 1
Optional Accessories
Cable and software for RS-232C
Customized testing range |
Description:
Operating principle: magnetic induction/eddy current (F/NF)
With separate probe(s)
Measuring range: 0-1250um/0-50mil
Resolution: 0.1/1
Accuracy: ±1-3%n or ±2.5um
Min. measuring area: 5mm
Min. sample thickness: 0.3mm
Battery indicator: low battery indicator
Metric/ imperial: convertible
Power supply: 4x1.5V AAA(UM-4)battery
Auto power off
Operating conditions: 0-+45?(32?-104?),=90%RH
Dimensions: 124x62x30mm
weight: 115g |
Coating Thickness Gauge CTG8828 (Designed for car industry)
|
Description:
In built probe(s)
Operating principle: magnetic induction/eddy current (F/NF)
Measuring range:0-1250um/0-50mil
Resolutio: 0.1/1
Accuracy: ±1-3%n or ±2.5um
Min. measuring area: 6mm
Min. sample thickness: 0.3mm
Battery indicator: low battery indicator
Metric/ imperial: convertible
Power supply: 4x1.5V AAA(UM-4)battery
Auto power off
Operating conditions:0-+45?(32?-104?),=90%RH
Dimensions: 126x65x27mm
weight: 81g(not including battery)
Optional accessories: other range 0-200um to 15000um
Standard Delivery of TG8828FN:
Main unit(1000um) 1
F type probe 1
N type probe 1
F calibration base set 1
N calibration base set 1
Calibration foil set 1set (4 foils)
Carrying case 1
Instruction manual 1
Optional Accessories
Cable and software for RS-232C
Customized testing range |
Coating Thickness Gauge CTG8826
|
Description:
With separate probe(s)
Function: TG-8826F (ferrous substrate) TG-8826N (non ferrous substrate) TG-8826FN (2 in 1, F & NF type)
Operating principle: magnetic induction/eddy current (F/NF)
Measuring range:0-1250um/0-50mil
Resolution; 0.1/1
Accuracy: ±1-3%n or ±2.5um
Min. measuring area: 6mm
Min. sample thickness: 0.3mm
Battery indicator: low battery indicator
Metric/ imperial: convertible
Power supply: 4x1.5V AAA(UM-4)battery
Auto power off
Operating conditions:0-+45?(32?-104?),=90%RH
Dimensions: 126x65x27mm
weight: 81g(not including battery)
Optional accessories: other range 0-200um to 15000um
Application:
1. TG-8826F can measure the thickness of non-magnetic coating layers (aluminum, chrome, copper, ename1, rubber, paint) on magnetic substrate (steel, iron, alloy and magnetic stainless steel)
2. TG-8826N can measure the thickness of non-conductive coating layers (ename1, rubber, paint, vanish, plastic anodic oxide layer) covered on non-ferrous substrate (aluminum, brass, zinc, tin and non-magnetic stainless steel)
Standard Delivery of TG8826FN:
Main unit(1000um) 1
F type probe 1
N type probe 1
F calibration base set 1
N calibration base set 1
Calibration foil set 1set (4 foils)
Carrying case 1
Instruction manual 1
Optional Accessories
Cable and software for RS-232C
Customized testing range |
Coating Thickness Gauge CTG8825
|
Description:
Function: TG-8825F (ferrous substrate) TG-8825N (non ferrous substrate) TG-8825FN (2 in 1, F & NF type)
Measuring principle: magnetic induction & eddy current
Measurement range:0~200um/500um /1000um/2000um in built probe (integrated probe)
Resolution:0.1 um (under 100um) 1um (over 100um)
Accuracy: ±1~3%n or ±3um
Probe type recognition: Manual (For TG-8825FN)
Measurement mode: single
Data output: RS232C interface (optional)
Power supply: 4x1.5V AAA(UM-4)battery
2 Shutdown models: manual / automatic
Dimension: 131x65x28mm
Application:
1. TG-8825F can measure the thickness of non-magnetic coating layers (aluminum, chrome, copper, ename1, rubber, paint) on magnetic substrate (steel, iron, alloy and magnetic stainless steel)
2. TG-8825N can measure the thickness of non-conductive coating layers (ename1, rubber, paint, vanish, plastic anodic oxide layer) covered on non-ferrous substrate (aluminum, brass, zinc, tin and non-magnetic stainless steel)
Standard Delivery of TG8825FN:
Main unit(1000um) 1
F type probe 1
N type probe 1
F calibration base set 1
N calibration base set 1
Calibration foil set 1set (4 foils)
Carrying case 1
Instruction manual 1
Optional Accessories